Application:
- Excellent to measure a magnetic flux of
air core coil.
- High frequency of use as MG-701AXL
wears better than the Standard Type Probe.
Note: Since the Axial Type Probe has a gap, about 0.5 mm approximately, between the magnetic flux sensor and an object to be measured, the measured value can be different from that of Standard Type Probe (Transverse type).
How to use:

1.
Plug the Axial Type Probe MG-701AXL
into the gauss meter.

2.
This is the way to measure an
axial magnetic field density whose magnetic direction is indicated on
the picture.
Contact vertically with the target object and consider the maximum value
as measured value.

3.
The sensor is positioned in
the center of the probe head as indicated on the picture. An accurate
value can not be measured out of the point.
To train on a target point, use the angle indexes as a guide, which are
marked on each 90 degree around the head.

4.
The sensor is not separable from the head.
Also do not overpower the head as indicated direction on the picture,
it can cause failure.